CALCulate<1|2>:MARKer<1 to 4>:Y:PERCent 0 to100%
This command positions the selected marker in the selected window to the given probability. If
marker 2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
Note: The command is only available with the CCDF measurement switched on.
The associated level value can be determined with the CALC:MARK:X? command.
Example: "CALC1:MARK:Y:PERC 95PCT" 'Positions marker 1 in screen A to a
'probability of 95%.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A
CALCulate<1|2>:MARKer<1 to 4>:MAXimum[:PEAK]
This command positions the marker to the current maximum value of the corresponding trace in the
selected measurement window. The corresponding marker is activated first or switched to the
marker mode.
Note: If no maximum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error eode: -200) is produced.
Example: "CALC:MARK2:MAX" 'Positions marker 2 in screen A to the maximum value of
'the trace.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1 to 4>:MAXimum:NEXT
This command positions the marker to the next smaller maximum value of the corresponding trace in
the selected measurement window.
Note: If no next smaller maximum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Example: "CALC:MARK2:MAX:NEXT" 'Positions marker 2 in screen A to the next
'lower maximum value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1 to 4>:MAXimum:RIGHt
This command positions the marker to the next smaller maximum value to the right of the current
value (i.e. in ascending X values) on the corresponding trace in the selected measurement window.
Note: If no next smaller maximum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Example: "CALC:MARK2:MAX:RIGH" 'Positions marker 2 in screen A to the next
'lower maximum value to the right of the
'current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM