Operating instructions

R&S ESCI CALCulate:MARKer Subsystem
1166.6004.12 6.45 E-1
CALCulate<1|2>:MARKer<1 to 4>:MAXimum:LEFT
This command positions the marker to the next smaller maximum value to the left of the current
value (i.e. in descending X values) on the trace in the selected measurement window.
Note: If no next smaller maximum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Example: "CALC:MARK2:MAX:LEFT" 'Positions marker 2 in screen A to the next
'lower maximum value to the left of the current
'value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1 to 4>:MINimum[:PEAK]
This command positions the marker to the current minimum value of the corresponding trace in the
selected measurement window. The corresponding marker is activated first or switched to marker
mode, if necessary.
Note: If no minimum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error code: -200) is produced.
Example: "CALC:MARK2:MIN" 'Positions marker 2 in screen A to the minimum value of
'the trace.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1 to 4>:MINimum:NEXT
This command positions the marker to the next higher minimum value of the corresponding trace in
the selected measurement window.
Note: If no next higher minimum value is found on the trace (level spacing to adjacent values <
peak excursion), an execution error (error code: -200) is produced.
Example: "CALC:MARK2:MIN:NEXT" 'Positions marker 2 in screen A to the next
'higher maximum value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.