3.1.6 Connecting an Optional DC Power Supply (R&S FPL1-
B30)
The R&S FPL1000 can also be equipped with an optional (internal) DC power
supply (R&S FPL1-B30). If installed, the R&S FPL1000 can be operated by a DC
voltage of 10.4 V to 28 V.
Risk of electric shock
Insufficient insulation and exceeding the current limitation of external power
supplies for safety extra-low DC voltage (SELV) can lead to electric shock.
Be sure to meet the requirements for reinforced/double insulation in accord-
ance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No. 1010.1)
orDIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). Provide current limi-
tation in accordance with DIN EN 61010-1 Appendix F2.1.
DC Connection
► Connect the DC power supply on the rear panel of the R&S FPL1000 to the
power source using the supplied power cable.
3.1.7 Connecting an Optional Battery Pack (R&S FPL1-B31)
As an alternative to the fixed AC or DC power supply, the R&S FPL1000 can be
operated by a battery pack if the option R&S FPL1-B31 is installed. If the battery
pack is installed and neither DC nor AC power is supplied, the R&S FPL1000
automatically switches to battery operation (during operation).
The R&S FPL1000 comes with the battery pack already installed if the battery
option was ordered together with the instrument. Otherwise, insert the battery
packs in the slots at the top of the rear panel.
Putting into Operation