User Manual

General Instrument Setup
R&S
®
FPL1000
527User Manual 1178.3370.02 ─ 09
The following units are available:
dB
dBm
dBmV
dBμV
dBμV/m
dBμA
dBμA/m
dBpW
dBpT
Compatibility
Indicates whether the transducer factors are compatible with the current measurement
settings.
For more information on which conditions a transducer line must fulfill to be compati-
ble, see Chapter 9.7.1, "Basics on Transducer Factors", on page 523.
Activating / Deactivating
Activates/deactivates the transducer line. Up to 8 transducer lines can be active at the
same time.
Remote command:
[SENSe:]CORRection:TRANsducer:SELect on page 961
[SENSe:]CORRection:TRANsducer[:STATe] on page 961
Comment
An optional description of the transducer line.
Included Transducer Lines in Overview (View Filter)
Defines which of the stored transducer lines are included in the overview. The view can
be restricted to compatible transducer lines only or include all transducer lines found.
Whether a line is compatible or not is indicated in the Compatibility setting.
Adjust Ref Level
Activates or deactivates the automatic adjustment of the reference level to the selected
transducer factor.
"Auto"
Activates the automatic adjustment. The original dynamic range is
restored by shifting the reference level by the maximum transducer
factor.
"Manual"
Deactivates the automatic adjustment. Adjust the reference level via
the "Amplitude" menu.
Remote command:
[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe] on page 959
Create New Line
Create a new transducer line.
Remote command:
[SENSe:]CORRection:TRANsducer:SELect on page 961
Transducers