General Instrument Setup
R&S
®
FPL1000
536User Manual 1178.3370.02 ─ 09
If the position of the instrument has been changed and you cannot look straight on
the screen
If another person operates the instrument
9.8.2 Alignment Settings
Access: [Setup] > "Alignment"
Both the instrument and the touchscreen can be aligned when necessary (see Chap-
ter 9.8.1, "Basics on Alignment", on page 534).
Start Self Alignment.................................................................................................... 536
Abort Self Alignment................................................................................................... 537
Await Warm-Up Operation before Self Alignment....................................................... 537
Shut down Device after Self Alignment.......................................................................537
Starting Touch Screen Alignment................................................................................537
Alignment Results:...................................................................................................... 537
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see Chapter 9.8.1, "Basics on Alignment", on page 534.
Note:
A running Sequencer operation is aborted when you start a self-alignment.
During self-alignment, do not connect a signal to the RF input connector. Running a
self-alignment with a signal connected to the RF input can lead to false measurement
results.
Remote command:
*CAL? on page 614, see also CALibration[:ALL] on page 956
Alignment