User Manual
Remote Commands
R&S
®
FPL1000
724User Manual 1178.3370.02 ─ 09
INIT;*WAI
//Initiates the measurement and finishes the sweep.
//-------------Retrieving the Results------------------------
CALC:MARK:FUNC:HARM:LIST?
//Queries the position of the harmonics.
CALC:MARK:FUNC:HARM:DIST? TOT
//Queries the total harmonic distortion.
11.6.12 Measuring the Third Order Intercept Point
● Determining the TOI..............................................................................................724
● Programming Example: Measuring the TOI..........................................................727
11.6.12.1 Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
●
CALCulate<n>:DELTamarker<m>:X on page 843
●
CALCulate<n>:DELTamarker<m>:X:RELative? on page 857
●
CALCulate<n>:DELTamarker<m>:Y on page 858
●
CALCulate<n>:MARKer<m>:X on page 845
●
CALCulate<n>:MARKer<m>:Y on page 858
Remote commands exclusive to TOI measurements
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe].......................................................... 724
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE...................................... 725
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?......................................................... 725
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult:MAXimum?......................................... 725
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult:MINimum?.......................................... 726
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe] <State>
This command initiates a measurement to determine the third intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instru-
ment. Marker 1 and marker 2 (both normal markers) are set to the maximum of the two
signals. Delta marker 3 and delta marker 4 are positioned to the intermodulation prod-
ucts. The delta markers can be modified separately afterwards with the
CALCulate<n>:DELTamarker<m>:X command.
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
Window
Configuring and Performing Measurements