Table 11-4: Optimization parameters in FFT mode
Optimization mode Description
DYNamic Optimizes the dynamic range by using the narrowest possible subspan (depending
on the RBW).
The autorange function for the internal IF gain calculation is activated to obtain the
best control range for the A/D converter.
SPEed Optimizes the sweep rate by using the widest possible subspan (depending on the
RBW).
The autorange function for the internal IF gain calculation is deactivated. (Note: set
the reference level accordingly to optimize the control range for the A/D converter).
It is recommended that you set the Sweep Time to "Auto" to optimize the sweep
rate.
AUTO Uses a medium-sized subspan to obtain a compromise between a large dynamic
range and a fast sweep rate.
The autorange function for the internal IF gain calculation is activated to obtain the
best control range for the A/D converter.
Zero span mode
For zero span measurements, the optimization mode defines the selection of the A/D
converter prefilter.
Table 11-5: Optimization parameters in zero span mode
Optimization mode Description
DYNamic The narrowest filter possible (depending on the RBW) is used.
SPEed The widest filter possible (depending on the RBW) is used.
AUTO A medium-sized prefilter is used.
Parameters:
<Mode> *RST: AUTO
Example:
SWE:OPT DYN
Selects optimization for dynamic range.
Manual operation: See "Optimization" on page 327
[SENSe:]SWEep[:WINDow<n>]:POINts <SweepPoints>
This command defines the number of sweep points to analyze after a sweep.
Note that the number of sweep points is limited to 10001 when measuring spurious
emissions.
Suffix:
<n>
.
Parameters:
<SweepPoints> Range: 101 to 100001
*RST: 1001
Example:
SWE:POIN 251