User Manual
Preparing for Use
R&S
®
FPL1000
16Getting Started 1323.1602.02 ─ 03
3.1.4 Connecting the AC Power
The R&S FPL1000 is equipped with an AC power supply connector. The
R&S FPL1000 can be used with different AC power voltages and adapts itself
automatically to it. Refer to the datasheet for the requirements of voltage and fre-
quency. The AC power connector is located on the rear panel of the instrument.
For details on the connector, refer to Chapter 4.2.2, "AC Power Supply Connec-
tion and Main Power Switch", on page 47.
► Connect the R&S FPL1000 to the AC power supply using the supplied power
cable.
Since the instrument is assembled in line with the specifications for safety
class EN61010, it must only be connected to an outlet that has a ground con-
tact.
3.1.5 Changing the AC Supply Fuse
If the battery option R&S FPL1-B31 is installed, an AC supply fuse is built into the
AC power supply connector. Only use fuses of the type 5A T IEC60127-2/V
(order no. 0099.6735.00). Such fuses are used for all of the specified nominal AC
supply voltages.
To change the AC supply fuses
1. Disconnect the power cable.
2. Open the flap covering the fuse holder using a small screwdriver (or similar).
3. Remove the fuse holder.
4. Remove the fuse and install the new one.
5. Reinsert the fuse holder.
6. Close the flap.