R&S®FPL1000 SIGNAL AND SPECTRUM ANALYZER Experience high performance wherever you take it Data Sheet Version 06.
AT A GLANCE The R&S®FPL1000 Teaser mit 1 Zeile Abstand signal and zurspectrum letzten Zeile analyzer oben makes measuring fast and simple. The intuitive touchscreen is straightforward and easy to use. With its solid RF performance, light weight and small footprint, Text 1 Zeile Abstand combines zum Teaserthe functionality of a benchtop instrument with the portability of a handheld the R&S®FPL1000 instrument. In an RF lab, the R&S®FPL1000 is as indispensable as an oscilloscope or multimeter.
BENEFITS Key facts Frequency range: 5 kHz to 7.
ONE INSTRUMENT FOR MULTIPLE APPLICATIONS The R&S®FPL1000 is a single measuring instrument for many types of measurements. You can use it for spectral measurements, for highly accurate power measurements with power sensors and for analyzing analog and digitally modulated signals. Spectrum analysis Signal analysis of analog and digitally modulated signals Even in its basic configuration, the R&S®FPL1000 is a true allrounder.
Scalar network analysis Noise figure and gain measurements Equipped with the R&S®FPL1-B9 option, the R&S®FPL1000 offers an internal CW source and a tracking generator for quick and easy measurement of frequency response, filters and attenuation. The n-dB down marker determines the 3 dB bandwidth of a bandpass filter at the press of a button. Precision is enhanced by through, short and open normalization methods.
SOLID RF PERFORMANCE Featuring a phase noise of –108 dBc (1 Hz) at 10 kHz offset (1 GHz carrier), a third-order intercept point of +20 dBm, 1 Hz to 10 MHz resolution bandwidth and –166 dBm displayed average noise level, the R&S®FPL1000 is comparable to higher class analyzers. This makes it the ideal tool for use in the lab, in production and for service tasks. The 1 dB attenuator step size (R&S®FPL1-B25 option) and the preamplifier (R&S®FPL1-B22 option) extend the usable dynamic range and sensitivity.
INTUITIVE USER INTERFACE Operating the R&S®FPL1000 is as intuitive as using a smartphone. You can configure the instrument and perform measurements with the touchscreen. A one-finger swipe across the screen adjusts the center frequency or the reference level. Two-finger gestures adjust the displayed span or level range. High-resolution display Flexible arrangement of results and MultiView The 10.1" screen with 1280 × 800 pixel resolution provides a precise representation of the signal.
10.1" HIGH-RESOLUTION DISPLAY 10.1" high-resolution display 1280 × 800 pixel resolution Hardkeys For setup, presets, etc. Two USB 2.
Soft menu selection ❙ Quick access to key tools ❙ Hardware settings at a glance Measurement setup buttons Control knob Numerical keypad With unit keys for frequency and level Generator output Internal generator output RF input Rohde & Schwarz R&S®FPL1000 Signal and Spectrum Analyzer 9
FULLY PORTABLE The R&S®FPL1000 signal and spectrum analyzer can be used almost everywhere. With a depth of only 23 cm, it fits into any workplace and leaves enough space for DUTs and other measuring instruments. Due to its low weight of 6 kg and the carrying handle, you can take it wherever you need it. Battery pack and 12 V/24 V power supply (option) The optional battery pack provides three hours of operation.
R&S®FPL1-K7 AM/FM/ΦM ANALOG DEMODULATION The R&S®FPL1-K7 option converts the R&S®FPL1000 into an analog modulation analyzer for amplitude, frequency and phase modulated signals. It measures the characteristics of the useful modulation and other items such as residual FM or synchronous modulation.
R&S®FPL1-K30 NOISE FIGURE AND GAIN MEASUREMENTS The R&S®FPL1-K30 noise figure and gain measurement option 1) allows you to characterize the most important amplifiers specifications. Using the Y-factor method, the noise figure and gain are measured with high accuracy independent of the instruments own noise figure. Typical applications for the R&S®FPL1-K30 include the characterization of amplifiers.
R&S®FPL1-K40 PHASE NOISE MEASUREMENT APPLICATION Phase noise is an important parameter in wireless communications systems. The R&S®FPL1-K40 option enables the R&S®FPL1000 to perform fast and easy phase noise measurements in development and production. Equipped with the R&S®FPL1-K40 option, the R&S®FPL1000 can measure single sideband phase noise across a selectable carrier offset frequency range displayed on a logarithmic axis.
R&S®FPL1-K54 EMI MEASUREMENT APPLICATION The R&S®FPL1-K54 EMI measurement application adds EMI diagnostic functionality to the R&S®FPL signal and spectrum analyzer. The R&S®FPL1-K54 offers EMI bandwidths for commercial and military applications, detectors such as quasi-peak, CISPR-average and RMS-average, limit lines and correction factors.
R&S®FPL1-K70 VECTOR SIGNAL ANALYSIS The R&S®FPL1000 analyzes and demodulates digitally modulated single-carrier signals with up to 40 MHz analysis bandwidth. Universities and research facilities will benefit from the instrument’s flexibility in analyzing proprietary signals. Developers of mobile communications devices and components can easily use the predefined standard settings.
R&S®VSE-K106 EUTRA/LTE NB-IOT MEASUREMENT SOFTWARE The R&S®FPL1000 can be used to analyze cellular 3GPP NB-IoT signals. It captures the signal, which is then analyzed by the R&S®VSE-K106 EUTRA/LTE NB-IoT measurement software 1). 1) R&S®VSE base software and R&S®FSPC license dongle required.
Specifications Frequency Frequency range Frequency resolution Scaling R&S®FPL1003 R&S®FPL1007 standard with R&S®FPL1-K54, RBW ≤ 1 MHz Reference frequency, internal, nominal Accuracy Aging per year Temperature drift (0 °C to +50 °C) Achievable initial calibration accuracy standard with R&S®FPL1-B4 OCXO reference frequency option standard with R&S®FPL1-B4 OCXO reference frequency option standard with R&S®FPL1-B4 OCXO reference frequency option Frequency readout Marker resolution Uncertainty Number of s
Spectral purity SSB phase noise frequency = 1000 MHz, carrier offset 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz 10 MHz nom. –88 dBc (1 Hz) < –99 dBc (1 Hz) < –105 dBc (1 Hz), typ. –108 dBc (1 Hz) < –110 dBc (1 Hz), typ. –115 dBc (1 Hz) < –130 dBc (1 Hz), typ. –135 dBc (1 Hz) nom.
Channel filters Bandwidths (–3 dB) 100/200/300/500 Hz 1/1.5/2/2.4/2.7/3/3.4/4/4.5/5/6/7.5/8.5/9/ 10/12.5/14/15/16/20/21/25/30/50/100/150/ 192/200/300/500 kHz 1/1.228/1.5/2/3/3.75/5/10 MHz nom. < 2 % nom. < 2 Bandwidth uncertainty Shape factor 60 dB:3 dB EMI filters (with R&S®FPL1-K54 option) Bandwidths (–6 dB) 10/100/200 Hz 1/9/10/100/120 kHz 1 MHz nom. < 3 % nom.
RF preamplifier on (gain nom.
< 0.2 dB (σ = 0.07 dB) Attenuator switching uncertainty f = 50 MHz, 0 dB to 45 dB, referenced to 10 dB attenuation Uncertainty of reference level setting Bandwidth switching uncertainty 0 dB 3 referenced to RBW = 10 kHz and sweep type FFT sweep type = FFT (RBW < 100 kHz) nom. < 0.1 dB sweep type = sweep (RBW ≥ 100 kHz) nom. < 0.2 dB Nonlinearity of displayed level Logarithmic level display Linear level display Total measurement uncertainty S/N > 16 dB, 0 dB to –50 dB S/N > 16 dB, 0 dB to –70 dB < 0.
Signal analysis bandwidth ≤ 10 MHz Amplitude flatness Deviation from linear phase Signal analysis bandwidth ≤ 40 MHz Amplitude flatness Deviation from linear phase fcenter ≥ 12 MHz and (1.25 × signal analysis bandwidth) fcenter ≥ 12 MHz and (1.25 × signal analysis bandwidth) nom. ±0.3 dB fcenter ≥ 12 MHz and (1.25 × signal analysis bandwidth) fcenter ≥ 12 MHz and (1.25 × signal analysis bandwidth) nom. ±0.5 dB nom. ±1° nom. ±1.
General data Display Resolution Pixel failure rate Data storage Internal External 21 cm LC TFT color display (10.1") 1280 × 800 pixel (WXGA resolution) < 1 × 10–5 standard solid-state drive (SSD) 32 Gbyte supports USB 2.0 compatible memory devices Climatic loading operating temperature range storage temperature range without condensation 0 °C to +50 °C –20 °C to +70 °C +40 °C at 85 % rel. humidity, in line with EN 60068-2-30, Mechanical resistance Vibration sinusoidal 5 Hz to 55 Hz 0.
Options R&S®FPL1-B5 additional interfaces User port Connector Output Input Noise source control and power sensor Connectors 25-pin D-Sub female TTL-compatible, 0 V/5 V, max. 15 mA TTL-compatible, max.
R&S®FPL1-B9 internal generator Frequency Frequency range Setting resolution R&S®FPL1003 R&S®FPL1007 independent CW source Frequency offset Setting range Setting resolution Spectral purity SSB phase noise Harmonics Non-harmonic spurious Level Specified level range Setting resolution Setting range Absolute level uncertainty Frequency response Level nonlinearity 8 9 0 Hz to fmax 8 0.
Maximum output power versus frequency, level in dBm (meas.). Minimum output power versus frequency, level in dBm (meas.). Dynamic range RBW = 1 kHz, f = 1 GHz nom. 115 dB Power sweep Specified level range Setting resolution Setting range –50 dBm to 0 dBm 0.1 dB –60 dBm to +10 dBm GEN output Connector VSWR N female, 50 Ω nom. 1.5 Reverse power DC voltage CW RF power Max.
R&S®FPL1-B30 DC power input 12 V/24 V Input voltage range DC Input current Vin = 12 V/24 V Vin = 12 V/24 V, operating mode, without internal batteries (R&S®FPL1-B31) Vin = 12 V/24 V, operating mode, internal batteries in charge mode Vin = 12 V/24 V, instrument standby mode, internal batteries in charge mode operating temperature range storage temperature range Temperature 12 V to 24 V (nom.), 10.4 V to 28 V, switch-on voltage > 11 V (meas.) 13 A/6.5 A (nom.) 5.5 A/2.7 A (meas.) 11 A/5 A (meas.) 6.
Ordering information Designation Signal and spectrum analyzer Signal and spectrum analyzer Accessories supplied Power cable and quick start guide Type R&S®FPL1003 R&S®FPL1007 Order No. 1304.0004.03 1304.0004.07 Options Designation OCXO reference frequency Additional interfaces Type R&S®FPL1-B4 R&S®FPL1-B5 Order No. 1323.1902.02 1323.1883.02 Retrofittable yes yes Internal generator Internal generator GPIB interface Second hard disk (SSD) R&S®FPL1-B9 R&S®FPL1-B9 R&S®FPL1-B10 R&S®FPL1-B19 1323.1925.
Designation Attenuator 50 W, 3/6/10/20/30 dB, 2 GHz Type R&S®RBU50 Attenuator 50 W, 20 dB, 6 GHz Connectors and cables IEC/IEEE bus cable, length: 1 m IEC/IEEE bus cable, length: 2 m DC block DC block, 10 kHz to 18 GHz (type N) R&S®RDL50 Order No. 1073.8695.xx (xx = 03/06/10/20/30) 1035.1700.52 R&S®PCK R&S®PCK 0292.2013.10 0292.2013.20 R&S®FSE-Z4 1084.7443.
Wideband power sensors 50 MHz to 18 GHz, 100 mW 50 MHz to 40 GHz, 100 mW (2.92 mm) 50 MHz to 40 GHz, 100 mW (2.40 mm) 50 MHz to 44 GHz, 100 mW (2.
Rohde & Schwarz R&S®FPL1000 Signal and Spectrum Analyzer
Rohde & Schwarz Worldwide Local und personalized Customized and flexible Uncompromising quality Long-term dependability The Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and measurement, broadcast and media, secure communications, cybersecurity, monitoring and network testing.