User Manual

Remote Commands
R&S
®
FPL1000
1005User Manual 1178.3370.02 ─ 09
Command Known Differences
FA Default value, range, step size
FB Default value, range, step size
ID
M2 Default value, range, step size
M3 Default value, range, step size
MKACT Only marker 1 is supported as the active marker.
MKBW Default value
MKPT Step size
MKPX Step size
OL? Storage of instrument settings:
80 characters are returned as information on the instrument settings.
The contents of the 80 characters returned does not correspond to the original data con-
tents of the 8566A/8568A family.
OL Readout of instrument settings:
The 80 characters read by means of OL? are accepted as information on the correspond-
ing instrument settings.
The contents of the 80 characters read does not correspond to the original data contents
of the 8566A/8568A family.
RB Default value, range, step size
RL Default value, step size
RLPOS Adapts the position of the reference level even if the tracking generator normalization is
not active.
RQS Supported bits:
1 (Units key pressed)
2 (End of Sweep)
3 (Device error)
4 (Command complete)
5 (Illegal command)
11.11.2.4 Model-Dependent Default Settings
If the GPIB language is switched over to an 85xx model, the GPIB address is automati-
cally switched over to 18 provided that the default address of the R&S FPL1000 (20) is
still set. If a different value is set, this value is maintained. Upon return to SCPI, this
address remains unchanged.
The following table shows the default settings obtained after a change of the GPIB lan-
guage and for the commands IP, KST and RESET:
Emulating Other Instruments' Commands