User Manual
Remote Commands
R&S
®
FPL1000
1011User Manual 1178.3370.02 ─ 09
STATus:PRESet
This command resets the edge detectors and ENABle parts of all registers to a defined
value. All PTRansition parts are set to FFFFh, i.e. all transitions from 0 to 1 are
detected. All NTRansition parts are set to 0, i.e. a transition from 1 to 0 in a
CONDition bit is not detected. The ENABle part of the STATus:OPERation and
STATus:QUEStionable registers are set to 0, i.e. all events in these registers are not
passed on.
Usage: Event
STATus:QUEue[:NEXT]?
This command queries the most recent error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
This command is identical to the SYSTem:ERRor[:NEXT]? command.
Usage:
Query only
11.12.2 Reading Out the CONDition Part
For more information on the condition part see Chapter 10.1.7.2, "Structure of a SCPI
Status Register", on page 561.
STATus:OPERation:CONDition?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:ACPLimit:CONDition? <ChannelName>
STATus:QUEStionable:EXTended:CONDition? <ChannelName>
STATus:QUEStionable:EXTended:INFO:CONDition? <ChannelName>
STATus:QUEStionable:FREQuency:CONDition? <ChannelName>
STATus:QUEStionable:LIMit<n>:CONDition? <ChannelName>
STATus:QUEStionable:LMARgin<n>:CONDition? <ChannelName>
STATus:QUEStionable:POWer:CONDition? <ChannelName>
STATus:QUEStionable:TEMPerature:CONDition? <ChannelName>
STATus:QUEStionable:TIME:CONDition? <ChannelName>
These commands read out the CONDition section of the status register.
The commands do not delete the contents of the CONDition section.
Suffix:
<n>
.
Window
Query parameters:
<ChannelName> String containing the name of the channel.
The parameter is optional. If you omit it, the command works for
the currently active channel.
Usage: Query only
Using the Status Register