User Manual

Remote Commands
R&S
®
FPL1000
1023User Manual 1178.3370.02 ─ 09
Suffix:
<n>
.
Window
<li> Limit line
<t> irrelevant
Parameters:
<TraceNumber> 1 to 6
*RST: 1
Example:
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
DISPlay[:WINDow<n>]:STATe
<State>
This command changes the display state of the selected measurement window.
Note that this command is maintained for compatibility reasons only. Use the LAYout
commands for new remote control programs
(See Chapter 11.7.1, "Working with Windows in the Display", on page 749).
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
DISPlay[:WINDow<n>]:TYPE <WindowType>
This command selects the results displayed in a measurement window.
Note that this command is maintained for compatibility reasons only. Use the LAYout
commands for new remote control programs (see Chapter 11.7.1, "Working with Win-
dows in the Display", on page 749).
The parameter values are the same as for LAYout:ADD[:WINDow]? on page 750.
Suffix:
<n>
.
Window
Parameters:
<WindowType> DIAGram | RSUMmary | MTABle | PEAKlist | SGRam
HCOPy:ITEM:ALL
This command is maintained for compatibility reasons only. It has no effect.
Deprecated Commands