User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
125User Manual 1178.3370.02 ─ 09
General power sweep settings are described in "Power Sweep Settings" on page 295.
Compression Measurement State...............................................................................125
Reference....................................................................................................................125
Compression Points....................................................................................................125
Compression Measurement State
Enables or disables the compression point measurement.
Remote command:
CONFigure:CMEasurement[:STATe] on page 637
Reference
Determines the reference point for the linear trace on which the deviation or compres-
sion calculation is based.
By default, the reference point is automatically set to the start level of the power sweep
(see "Level Start / Level Stop" on page 297). In manual mode, define the reference
point as required.
For details, see "Reference point" on page 122.
Remote command:
CONFigure:CMEasurement:REFerence[:MODE] on page 635
CONFigure:CMEasurement:REFerence:VALue on page 634
Compression Points
Selects which compression points to calculate. For the user-defined compression
point, define the distance from the linear trace to the reference trace to base the calcu-
lation on. For details, see Chapter 7.2.3.1, "About the Power Sweep Compression
Point Measurement", on page 121.
Measurements and Results