User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
175User Manual 1178.3370.02 ─ 09
RF Attenuation............................................................................................................ 176
Preamp........................................................................................................................177
Transducer Factor.......................................................................................................177
Limit Check <n>.......................................................................................................... 177
Abs Limit Start / Stop <n>........................................................................................... 177
Rel Limit Start / Stop <n>............................................................................................177
Multi-Limit Calc <n>.................................................................................................... 178
Min Sweep Points....................................................................................................... 178
Insert before Range / Insert after Range.....................................................................179
Delete Range.............................................................................................................. 179
Symmetrical Setup......................................................................................................179
Range Start / Range Stop
Sets the start frequency/stop frequency of the selected range.
To change the start/stop frequency of the first or last range, respectively, select the
appropriate span in the [SPAN] configuration dialog. You can set a span that is smaller
than the overall span of the ranges. In this case, the measurement includes only the
ranges that lie within the defined span and have a minimum span of 20 Hz. The first
and last ranges are adapted to the given span as long as the minimum span of 20 Hz
is not violated.
Define frequency values for each range relative to the center frequency. Center the ref-
erence range on the center frequency. The current "Tx Bandwidth" defines the mini-
mum span of the reference range (see "Channel Power Settings" on page 181).
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STARt on page 666
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STOP on page 666
Fast SEM
Activates "Fast SEM" mode for all ranges in the sweep list. For details, see "Fast SEM
Measurements" on page 168.
Note: If you deactivate "Fast SEM" mode while "Symmetrical Setup" mode is on,
"Symmetrical Setup" mode is automatically also deactivated.
If you activate "Fast SEM" mode while "Symmetrical Setup" mode is on, not all range
settings can be configured symmetrically automatically.
Remote command:
[SENSe:]ESPectrum<sb>:HSPeed on page 663
Filter Type
Sets the filter type for this range.
For details on filter types, see Chapter 7.6.1.6, "Which Data May Pass: Filter Types",
on page 320.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:FILTer:TYPE on page 665
RBW
Sets the resolution bandwidth for this range.
Measurements and Results