User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
176User Manual 1178.3370.02 ─ 09
For details on the RBW, see Chapter 7.6.1.1, "Separating Signals by Selecting an
Appropriate Resolution Bandwidth", on page 317.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:RESolution on page 664
VBW
Sets the video bandwidth for this range.
For details on the VBW, see Chapter 7.6.1.2, "Smoothing the Trace Using the Video
Bandwidth", on page 318.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:VIDeo on page 664
Sweep Time Mode
Activates or deactivates the auto mode for the sweep time.
Currently, only auto mode is available for the R&S FPL1000.
For details on the sweep time mode, see Chapter 7.6.1.7, "How Long the Data is Mea-
sured: Sweep Time", on page 320
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME:AUTO on page 675
Sweep Time
Sets the sweep time value for the range.
For details on the sweep time, see Chapter 7.6.1.7, "How Long the Data is Measured:
Sweep Time", on page 320
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME on page 675
Ref Level
Sets the reference level for the range.
For details on the reference level, see Chapter 7.5.1.1, "Reference Level",
on page 308.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:RLEVel on page 674
RF Att Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see Chapter 7.5.1.2, "RF Attenuation", on page 309.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation:AUTO on page 667
RF Attenuation
Sets the attenuation value for the range.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation on page 667
Measurements and Results