The Spectrum Application (RF Measurements)
R&S
®
FPL1000
180User Manual 1178.3370.02 ─ 09
Sub Block Count......................................................................................................... 180
Sub Block / Center Freq..............................................................................................180
Standard......................................................................................................................180
Edit Sweep List........................................................................................................... 180
Sub Block Count
Defines the number of sub blocks. By default, the familiar SEM measurement with just
one single block of ranges is configured.
Remote command:
[SENSe:]ESPectrum<sb>:SCOunt on page 662
Sub Block / Center Freq
Defines the center frequency for an individual sub block. The center frequency deter-
mines the reference range used for each block.
For measurements with only one sub block, this setting corresponds to the global set-
ting in the "Frequency" settings (see Center Frequency).
Remote command:
[SENSe:]ESPectrum<sb>:SCENter on page 662
Standard
Defines the use of a standard settings file for a particular sub block. For details, see
"Standard Files" on page 183.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet[:STANdard] on page 660
Edit Sweep List
Switches to the "Sweep List" tab of the "Spectrum Emission Mask" dialog box to con-
figure the individual frequency ranges and mask limits for the corresponding sub block.
See "Sweep List" on page 174.
Reference Range
Access: "Overview" > "Select Measurement" > "Spectrum Emission Mask" > "Refer-
ence Range"
The range around the center frequency is defined as the reference range for all other
ranges in the sweep list.
Measurements and Results