User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
181User Manual 1178.3370.02 ─ 09
Power Reference Type................................................................................................181
Channel Power Settings..............................................................................................181
Tx Bandwidth................................................................................................ 181
RRC Filter State............................................................................................181
Alpha:............................................................................................................182
Power Reference Type
Defines how the reference power is calculated.
"Channel Power"
Measures the channel power within the reference range using the
integration bandwidth method. Additional settings can be configured
for this method.
"Peak Power"
Determines the peak power within the reference range.
Remote command:
[SENSe:]ESPectrum<sb>:RTYPe on page 678
Channel Power Settings
If the "Power Reference Type:" "Channel Power" was selected, additional parameters
can be configured.
Tx Bandwidth ← Channel Power Settings
Defines the bandwidth used for measuring the channel power, with:
Minimum span ≤ "Tx Bandwidth" ≤ of reference range
Remote command:
[SENSe:]ESPectrum<sb>:BWID on page 677
RRC Filter State ← Channel Power Settings
Activates or deactivates the use of an RRC filter.
Measurements and Results