The Spectrum Application (RF Measurements)
R&S
®
FPL1000
182User Manual 1178.3370.02 ─ 09
Remote command:
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe] on page 677
Alpha: ← Channel Power Settings
Sets the alpha value of the RRC filter (if activated).
Remote command:
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa on page 677
Power Classes
Access: "Overview" > "Select Measurement" > "Spectrum Emission Mask" > "Power
Classes"
You can configure power classes which you can then assign to sweep list ranges. For
details, see "Power classes" on page 165.
Used Power Classes:..................................................................................................182
PMin/ PMax.................................................................................................................183
Sweep List...................................................................................................................183
Adding or Removing a Power Class........................................................................... 183
Used Power Classes:
Defines which power classes are considered for the SEM measurement. Limits can be
defined only for used power classes. It is only possible to select either one specific
power class or all the defined power classes.
If "All" is selected, the power class that corresponds to the currently measured power
in the reference range is used for monitoring. The limits assigned to that power class
are applied (see "Abs Limit Start / Stop <n>" on page 177 and "Rel Limit Start / Stop
<n>" on page 177).
Measurements and Results