User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
191User Manual 1178.3370.02 ─ 09
1. Configure and perform an SEM measurement as described in Chapter 7.2.7.6,
"How to Perform a Spectrum Emission Mask Measurement", on page 187.
2. In the "Overview", select the "Analysis" button.
3. If necessary, change the "Decimal Separator" to "COMMA" for evaluation in other
languages.
4. Select the "Save" button.
5. In the file selection dialog box, select a storage location and filename for the result
file.
6. Select the "Save" button.
The file with the specified name and the extension .dat is stored in the defined
storage location.
7.2.7.7 Measurement Example: Multi-SEM Measurement
The following measurement example demonstrates an SEM measurement for a signal
with multiple sub blocks.
A programming example demonstrating a SEM measurement in a remote environment
is provided in Chapter 11.6.7.10, "Example: SEM Measurement", on page 685.
Test setup:
Signal generator settings (e.g. R&S FPL1000 SMW):
Device Standard Center frequency Level Test model
SigGen 1 3GPP/FDD 900 MHz 0 dBm 1-16
SigGen 2 EUTRA/LTE 906.5 MHz 0 dBm 1_1_5MHz
Measurements and Results