The Spectrum Application (RF Measurements)
R&S
®
FPL1000
206User Manual 1178.3370.02 ─ 09
Remote command:
[SENSe:]LIST:RANGe<ri>:SWEep:TIME on page 695
Detector
Sets the detector for the range.
For details, refer to "Mapping Samples to sweep Points with the Trace Detector"
on page 386.
Remote command:
[SENSe:]LIST:RANGe<ri>:DETector on page 691
Reference Level
Sets the reference level for the range.
For details on the reference level, see Chapter 7.5.1.1, "Reference Level",
on page 308.
Remote command:
[SENSe:]LIST:RANGe<ri>:RLEVel on page 695
RF Attenuation Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see Chapter 7.5.1.2, "RF Attenuation", on page 309.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO on page 693
RF Attenuation
Sets the attenuation value for that range.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation on page 693
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see "Preamplifier" on page 313.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe on page 693
Sweep Points
Sets the number of sweep points for the specified range.
For details on sweep points, see Chapter 7.6.1.8, "How Much Data is Measured:
Sweep Points and Sweep Count", on page 321.
Remote command:
[SENSe:]LIST:RANGe<ri>:POINts[:VALue] on page 695
Stop After Sweep
This function is not available for the R&S FPL1000.
Measurements and Results