Example:
Linear axis:
With a linear axis, the distance between the sweep points is equal, e.g. 200 kHz.
Logarithmic axis:
With a logarithmic axis, the distance between sweep points is variable. In the spectrum
from 10 Hz to 100 Hz, the distance is a few Hz. Between 100 MHz and 1 GHz, the dis-
tance is several MHz.
The R&S FPL1000 EMI measurement supports a maximum of 200001 sweep points.
This number is based on typical bands measured with a single resolution bandwidth.
There are sufficient sweep points to make sure that a signal is found during the refined
measurement, even when covering 30 MHz to 1 GHz with logarithmic scaling and
120 kHz RBW.
Controlling V-Networks (LISN)
For measurements on power lines, the R&S FPL1 EMI measurement adds functionality
to control a line impedance stabilization network (LISN) directly. Thus you can deter-
mine the interference caused by power supplies and cables.
This feature requires the optional additional interfaces (R&S FPL1-B5).
You can connect the LISN to the user port of the R&S FPL1000. Control cables for the
various LISNs are available as accessories. The R&S FPL1000 then controls which
phase of the LISN is to be tested and outputs the information to the user port.
The R&S FPL1 EMI measurement supports several V-networks. For each type of net-
work, you can define the phase you want to test for interferences. The R&S FPL1 EMI
measurement allows you to test one phase at a time.
Table 7-19: Supported networks and phases
Network type Phases
Two-line V-networks
ESH3-Z5 N, L1
ENV216 N, L1
Four-line V-networks
ESH2-Z5 N, L1. L2, L3
ENV4200 N, L1. L2, L3
ENV432 N, L1. L2, L3