The Spectrum Application (RF Measurements)
R&S
®
FPL1000
280User Manual 1178.3370.02 ─ 09
Remote command:
[SENSe:]PMETer<p>:FREQuency:LINK on page 796
Unit/Scale
Selects the unit with which the measured power is to be displayed. Available units are
dBm, dB, W and %.
If dB or % is selected, the display is relative to the reference value that is defined with
either the "Meas -> Ref" setting or the "Reference Value" setting.
Remote command:
UNIT<n>:PMETer<p>:POWer on page 799
UNIT<n>:PMETer<p>:POWer:RATio on page 799
Meas Time/Average
Selects the measurement time or switches to manual averaging mode. In general,
results are more precise with longer measurement times. The following settings are
recommended for different signal types to obtain stable and precise results:
"Short"
Stationary signals with high power (> -40dBm), because they require
only a short measurement time and short measurement time provides
the highest repetition rates.
"Normal"
Signals with lower power or modulated signals
"Long"
Signals at the lower end of the measurement range (<-50 dBm) or
Signals with lower power to minimize the influence of noise
"Manual"
Manual averaging mode. The average count is set with the Average
Count (Number of Readings) setting.
Remote command:
[SENSe:]PMETer<p>:MTIMe on page 796
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe] on page 797
Setting the Reference Level from the Measurement Meas -> Ref
Sets the currently measured power as a reference value for the relative display. The
reference value can also be set manually via the Reference Value setting.
Remote command:
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE on page 794
Reference Value
Defines the reference value in dBm used for relative power meter measurements.
Remote command:
CALCulate<n>:PMETer<p>:RELative[:MAGNitude] on page 794
Use Ref Level Offset
If activated, takes the reference level offset defined for the analyzer into account for the
measured power (see "Shifting the Display (Offset)" on page 311).
If deactivated, takes the Sensor Level Offset into account.
Remote command:
[SENSe:]PMETer<p>:ROFFset[:STATe] on page 798