Overloading
At a reference level of -13 dBm and at an internal generator output level of the same
value, the R&S FPL1000 operates without overrange reserve. That means the
R&S FPL1000 is in danger of being overloaded if a signal is applied whose amplitude
is higher than the reference line. In this case, an overload message is displayed in the
status line.
Overloading can be avoided as follows:
Reducing the output level of the internal generator ("Level" on page 291)
Increasing the reference level (Reference Level in the "Amplitude" menu)
7.3.3.2 Internal Generator Settings
Access: Toolbar > "Generator Config"
Or: [Freq] > "Generator Config"
The internal generator signal is only available if the optional Internal Generator
R&S FPL1-B9 is installed.
The preview in the dialog box visualizes the configured signal from the internal genera-
tor.
For details, see Chapter 7.3.3.1, "Basics on the Internal Generator", on page 283.
Independent CW Source Settings
Access: Toolbar > "Generator Config"
Receiving Data Input and Providing Data Output