User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
397User Manual 1178.3370.02 ─ 09
Function Trace Settings
Trace 3: Min Hold
Auto Detector (Negative Peak)
Traces 4-6: Blank
Auto Detector
Trace 1/ Trace 2/ Trace 3/ Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
For details see Chapter 7.10.1.3, "How to Configure a Standard Trace", on page 397.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe] on page 817
Copy Trace
Access: "Overview" > "Analysis" > "Traces" > "Copy Trace"
Or: [TRACE] > "Copy Trace"
Copies trace data to another trace.
The first group of buttons (labeled "Trace 1" to "Trace 6") selects the source trace. The
second group of buttons (labeled "Copy to Trace 1" to "Copy to Tace 6") selects the
destination.
Remote command:
TRACe<n>:COPY on page 821
7.10.1.3 How to Configure a Standard Trace
Step-by-step instructions on configuring the trace settings are provided here.
For details on individual functions and settings see Chapter 7.10.1.2, "Trace Settings",
on page 393.
The remote commands required to perform these tasks are described in Chap-
ter 11.8.7, "Configuring the Trace Display and Retrieving Trace Data", on page 815.
Trace settings are configured in the "Traces" dialog box.
To display the "Traces" dialog box, do one of the following:
●
Press the [TRACE] key and then select the "Trace Config" softkey.
●
Select "Analysis" from the "Overview", then select the "Traces" tab.
1. For each trace, select the "Trace Mode" and "Trace Detector". Traces with the
trace mode "Blank" are not displayed.
2. To configure several traces to predefined display modes in one step, press the but-
ton for the required function:
● "Preset All Traces"
● "Set Trace Mode Max | Avg | Min"
● "Set Trace Mode Max | ClrWrite | Min"
For details see Chapter 7.10.1.2, "Trace Settings", on page 393.
Trace Configuration