User Manual
Getting Started
R&S
®
FPL1000
41User Manual 1178.3370.02 ─ 09
Pin Signal Description
7 GND Ground
8 to 11
not used for spectrum analysis
12 GND Ground
13 +5 V / max. 250 mA Supply voltage for external circuits
14 to 19 I/O_<no.> Control lines for user ports (see user manual)
20 GND Ground
21 READY FOR TRIG-
GER
Signal indicating that the instrument is ready to receive a trigger
signal.
The signal polarity is configurable.
22 to 25
not used for spectrum analysis
23 GND Ground
24 to 25
not used for spectrum analysis
5.2.2.11 IF/Video Output
The female BNC connector can be used for various outputs in the Spectrum applica-
tion:
●
Intermediate frequency (IF) output of approximately 20 MHz
●
Video output (1V)
Which output is provided is defined in the software ("Overview" > "Output").
This connector is only available if the R&S FPL1-B5 option is installed.
For details, see "Data Output" on page 300.
5.2.2.12 Headphones Connector
The R&S FPL1000 provides demodulators for AM, FM and PM signals, which can be
routed to the headphone connector. With headphones or an external loudspeaker con-
nected to the 3.5 mm headphone socket, the displayed signal can be identified acousti-
cally.
This connector is only available if the R&S FPL1-B5 option is installed.
Note the safety information provided in "Connecting headphones" on page 11.
For details, see "Data Output" on page 300.
5.2.2.13 Noise Source Control
The Noise Source Control female connector is used to provide the supply voltage for
an external noise source. For example, use it to measure the noise figure and gain of
amplifiers and frequency converting devices.