User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
428User Manual 1178.3370.02 ─ 09
This setting does not have any effect on limit lines that are defined by absolute values
for the x-axis.
Remote command:
CALCulate<n>:LIMit<li>:CONTrol:OFFSet on page 903
Y-Offset
Shifts a limit line that has relative values for the y-axis (levels or linear units such as
volt) vertically.
This setting does not have any effect on limit lines that are defined by absolute values
for the y-axis.
Remote command:
CALCulate<n>:LIMit<li>:LOWer:OFFSet on page 905
CALCulate<n>:LIMit<li>:UPPer:OFFSet on page 909
Create New Line
Creates a new limit line.
Edit Line
Edit an existing limit line configuration.
Copy Line
Copy the selected limit line configuration to create a new line.
Remote command:
CALCulate<n>:LIMit<li>:COPY on page 911
Delete Line
Delete the selected limit line configuration.
Remote command:
CALCulate<n>:LIMit<li>:DELete on page 912
Disable All Lines
Disable all limit lines in one step.
Remote command:
CALCulate<n>:LIMit<li>:STATe on page 912
Limit Line Details
Access: "Overview" > "Analysis" > "Lines" > "Limit Lines" > "New" / "Edit" / "Copy To"
or: [LINES] > "Line Config" > "Limit Lines" > "New" / "Edit" / "Copy To"
Display and Limit Lines