User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
432User Manual 1178.3370.02 ─ 09
For details on the file format see Chapter 7.11.2.4, "Reference: Limit Line File Format",
on page 436.
The limit line can be imported again later by the R&S FPL1000 for use in other mea-
surements.
Remote command:
MMEMory:STORe<n>:LIMit on page 913
File Explorer ← Export
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
7.11.2.3 How to Define Limit Lines
Access: "Overview" > "Analysis" > "Lines" > "Limit Lines"
or: [LINES] > "Line Config" > "Limit Lines"
Limit lines for spurious and SEM measurements
Note that for spurious and SEM measurements, special limit lines can be defined for
each frequency range, see "Limit Lines in SEM Measurements" on page 166 and "Limit
Lines in Spurious Measurements" on page 202. It is strongly recommended that you
define limits only via the "Sweep List" dialog for these measurements, not using the
[Lines] key.
Any changes to the special limit lines are automatically overwritten when the sweep list
settings are changed.
The following tasks are described here:
●
"How to find compatible limit lines" on page 432
●
"How to activate and deactivate a limit check" on page 433
●
"How to edit existing limit lines" on page 433
●
"How to copy an existing limit line" on page 433
●
"How to delete an existing limit line" on page 433
●
"How to configure a new limit line" on page 434
●
"How to move the limit line vertically or horizontally" on page 435
How to find compatible limit lines
► In the "Line Config" dialog box, select the "View Filter" option: "Show Compatible".
All stored limit lines with the file extension .LIN in the limits subfolder of the
main installation folder of the instrument that are compatible to the current mea-
surement settings are displayed in the overview.
Display and Limit Lines