User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
452User Manual 1178.3370.02 ─ 09
Generally, the format of this ASCII file can be processed by spreadsheet calculation
programs, e.g. MS-Excel. Different language versions of evaluation programs may
require a different handling of the decimal point. Thus you can define the decimal sep-
arator to be used (decimal point or comma, see "Decimal Separator" on page 448).
If the spectrogram display is selected when you select the "ASCII Trace Export" soft-
key, the entire histogram buffer with all frames is exported to a file. The data corre-
sponding to a particular frame begins with information about the frame number and the
time that frame was recorded.
Table 7-30: ASCII file format for trace export in the Spectrum application
File contents Description
Header data
Type;R&S FPL1000; Instrument model
Version;1.00; Firmware version
Date;01.Oct 2006; Date of data set storage
Mode;ANALYZER; Operating mode
Preamplifier;OFF Preamplifier status
Transducer; OFF Transducer status
Center Freq;55000;Hz Center frequency
Freq Offset;0;Hz Frequency offset
Start;10000;Hz
Stop;100000;Hz
Start/stop of the display range.
Unit: Hz for span > 0, s for span = 0, dBm/dB for statistics mea-
surements
Span;90000;Hz Frequency range (0 Hz in zero span and statistics measure-
ments)
Ref Level;-30;dBm Reference level
Level Offset;0;dB Level offset
Rf Att;20;dB Input attenuation
El Att;2.0;dB Electrical attenuation
RBW;100000;Hz Resolution bandwidth
VBW;30000;Hz Video bandwidth
SWT;0.005;s Sweep time
Sweep Count;20; Number of sweeps set
Ref Position;75;% Position of reference level referred to diagram limits (0 % =
lower edge)
Level Range;100;dB Display range in y direction. Unit: dB with x-axis LOG, % with x-
axis LIN
x-Axis;LIN; Scaling of x-axis linear (LIN) or logarithmic (LOG)
y-Axis;LOG; Scaling of y-axis linear (LIN) or logarithmic (LOG)
Importing and Exporting Measurement Results for Evaluation