User Manual
Getting Started
R&S
®
FPL1000
48User Manual 1178.3370.02 ─ 09
In the "Frequency" dialog box, set the center frequency to 50
MHz and tap "Zero
Span".
6. Increase the reference level so you can see the level of the calibration signal at
1 dBm.
a) In the "Amplitude" dialog box, set the "Reference Level" value to +10
dBm.
As the calibration signal does not vary over time, the level versus time diagram dis-
plays a straight line.
Figure 5-10: Time domain display of the calibration signal
7. Create a new measurement channel for I/Q analysis:
a) Press the [Mode] key.
b) Tap the "IQ Analyzer" button to activate a measurement channel for the I/Q
Analyzer application.
c) Tap the "Display Config" softkey to activate the SmartGrid mode.
Trying Out the Instrument