User Manual
Network and Remote Operation
R&S
®
FPL1000
568User Manual 1178.3370.02 ─ 09
Bit No. Meaning
4 ALT2 UPPer FAIL
This bit is set if the limit is exceeded in the upper 2nd alternate channel.
5 ALT2 LOWer FAIL
This bit is set if the limit is exceeded in the lower 2nd alternate channel.
6 ALT3 … 11 LOWer/UPPer FAIL
This bit is set if the limit is exceeded in one of the lower or upper alternate channels 3 … 11.
7 CACLR FAIL
This bit is set if the CACLR limit is exceeded in one of the gap channels.
8-14 Unused
15 This bit is always 0.
STATus:QUEStionable:EXTended Register
The STATus:QUEStionable:EXTended register contains further status information
not covered by the other status registers of the R&S FPL1000. A separate EXTended
register exists for each active channel.
You can read out the register with STATus:QUEStionable:EXTended:CONDition?
or STATus:QUEStionable:EXTended[:EVENt]?
Table 10-11: Meaning of the bits used in the STATus:QUEStionable:EXTended register
Bit No. Meaning
0 not used
1 INFO
This bit is set if a status message is available for the application.
Which type of message occurred is indicated in the STATus:QUEStionable:EXTended:INFO
Register.
2 to 14 Unused
15 This bit is always 0.
STATus:QUEStionable:EXTended:INFO Register
The STATus:QUEStionable:EXTended:INFO register contains information on the
type of messages that occur during operation of the R&S FPL1000. A separate INFO
register exists for each active channel.
You can read out the register with STATus:QUEStionable:EXTended:INFO:
CONDition? or STATus:QUEStionable:EXTended:INFO[:EVENt]?. You can
query all messages that occur for a specific channel using the command SYSTem:
ERRor:EXTended? on page 973.
Remote Control Basics