User Manual

Network and Remote Operation
R&S
®
FPL1000
585User Manual 1178.3370.02 ─ 09
Language.................................................................................................................... 585
IF Gain........................................................................................................................ 585
Sweep Repeat.............................................................................................................585
Coupling......................................................................................................................586
Revision String............................................................................................................586
Resetting the Factory Revision................................................................................... 586
Language
Defines the system language used to control the instrument.
For details on the available GPIB languages, see Chapter 11.11.2, "Reference: GPIB
Commands of Emulated HP Models", on page 980.
Note: Emulating previous R&S signal and spectrum analyzers. This function is also
used to emulate previous R&S signal and spectrum analyzers.
As a rule, the R&S FPL1000 supports most commands from previous R&S signal and
spectrum analyzers such as the FSQ, FSP, FSU, or FSV. However, the default values,
in particular the number of sweep points or particular bandwidths, may vary. Therefore,
the R&S FPL1000 can emulate these other devices, including their default values, in
order to repeat previous measurements or support existing control applications as in
legacy systems.
Remote command:
SYSTem:LANGuage on page 978
IF Gain
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz.
NORM Optimized for high dynamic range, overload limit is close to reference level.
PULS Optimized for pulsed signals, overload limit up to 10 dB above reference level.
This setting is only available if an HP language is selected (see "Language"
on page 585).
Remote command:
SYSTem:IFGain:MODE on page 978
Sweep Repeat
Controls a repeated sweep of the E1 and MKPK HI HP model commands (for details
on the commands refer to Chapter 11.11.2, "Reference: GPIB Commands of Emulated
HP Models", on page 980). If the repeated sweep is OFF, the marker is set without
sweeping before.
Note: In single sweep mode, switch off this setting before you set the marker via the
E1 and MKPK HI commands in order to avoid sweeping again.
This setting is only available if a HP language is selected (see "Language"
on page 585).
Remote command:
SYSTem:RSWeep on page 979
Network and Remote Control Settings