User Manual
Network and Remote Operation
R&S
®
FPL1000
586User Manual 1178.3370.02 ─ 09
Coupling
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW)
●
resolution bandwidth and video bandwidth (RBW/VBW)
For FSx, the standard parameter coupling of the instrument is used. As a result, in
most cases a shorter sweep time is used than in case of HP.
This setting is only available if a HP language is selected (see "Language"
on page 585).
Remote command:
SYSTem:HPCoupling on page 977
Revision String
Defines the response to the REV? query for the revision number.
(HP emulation only, see "Language" on page 585).
Max. 36 characters are allowed.
Remote command:
SYSTem:REVision[:STRing] on page 979
Resetting the Factory Revision
Resets the response to the REV? query for the revision number to the factory default
(HP emulation only, see "Language" on page 585).
Remote command:
SYSTem:REVision:FACTory on page 971
10.3.4 LXI Settings
Access: [SETUP] > "Network + Remote" > "LXI" tab
On the R&S FPL1000 the LXI Class C functionality is already installed and enabled;
thus, the instrument can be accessed via any web browser (e.g. the Microsoft Internet
Explorer) to perform the following tasks:
●
modifying network configurations
●
modifying device configurations
●
monitoring connections from the device to other devices
The "LXI" tab of the "Network + Remote" dialog box provides basic LXI functions and
information for the R&S FPL1000.
Alternatively, you can change the LAN settings using the LXI Web browser interface.
For details see Chapter 10.4.1.4, "How to Configure the LAN Using the LXI Web
Browser Interface", on page 595.
Only user accounts with administrator rights are able to use LXI functionality.
Network and Remote Control Settings