User Manual
Remote Commands
R&S
®
FPL1000
615User Manual 1178.3370.02 ─ 09
*IDN?
Identification
Returns the instrument identification.
Usage: Query only
*IST?
Individual status query
Returns the contents of the IST flag in decimal form. The IST flag is the status bit
which is sent during a parallel poll.
Return values:
<ISTflag> 0 | 1
Usage: Query only
*OPC
Operation complete
Sets bit 0 in the event status register when all preceding commands have been execu-
ted. This bit can be used to initiate a service request. The query writes a "1" into the
output buffer when all preceding commands have been executed, which is useful for
command synchronization.
*OPT?
Option identification query
Queries the options included in the instrument. For a list of all available options and
their description, refer to the data sheet.
Usage:
Query only
*PCB <Address>
Pass control back
Indicates the controller address to which remote control is returned after termination of
the triggered action.
Setting parameters:
<Address> Range: 0 to 30
Usage: Setting only
*PRE <Value>
Parallel poll register enable
Common Commands