*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the TRIGger
subsystem.
*TRG corresponds to the INITiate:IMMediate command (see INITiate<n>[:
IMMediate] on page 629).
Usage: Event
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code.
Note: If you start a self-test remotely, then select the "Local" softkey while the test is
still running, the instrument only returns to the manual operation state after the test is
completed. In this case, the self-test cannot be aborted.
Return values:
<ErrorCode> integer > 0 (in decimal format)
An error occurred.
0
No errors occurred.
Usage: Query only
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage:
Event
11.4 Commands for Remote Instrument Operation
The following commands are required to shutdown or reboot the R&S FPL1000 from a
remote PC.
SYSTem:CLOGging....................................................................................................... 618
SYSTem:REBoot............................................................................................................618
SYSTem:SHUTdown...................................................................................................... 618
Commands for Remote Instrument Operation