User Manual
Remote Commands
R&S
®
FPL1000
619User Manual 1178.3370.02 ─ 09
DISPlay:ATAB <State>
This command switches between the MultiView tab and the most recently displayed
channel setup. If only one channel setup is active, this command has no effect.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
INSTrument:CREate:DUPLicate
This command duplicates the currently selected channel setup, i.e creates a new chan-
nel setup of the same type and with the identical measurement settings. The name of
the new channel setup is the same as the copied channel setup, extended by a con-
secutive number (e.g. "IQAnalyzer" -> "IQAnalyzer 2").
The channel setup to be duplicated must be selected first using the INST:SEL com-
mand.
Example:
INST:SEL 'IQAnalyzer'
INST:CRE:DUPL
Duplicates the channel setup named 'IQAnalyzer' and creates a
new channel setup named 'IQAnalyzer2'.
Usage: Event
Manual operation: See "Duplicate Current Channel" on page 83
INSTrument:CREate[:NEW] <ChannelType>, <ChannelName>
This command adds an additional measurement channel. You can configure up to 10
measurement channels at the same time (depending on available memory).
See also
●
INSTrument[:SELect] on page 622
●
INSTrument:DELete on page 620
Parameters:
<ChannelType> Channel type of the new channel.
For a list of available channel types see INSTrument:LIST?
on page 620.
<ChannelName> String containing the name of the channel.
Note that you can not assign an existing channel name to a new
channel; this will cause an error.
Example:
INST:CRE SAN, 'Spectrum 2'
Adds an additional spectrum display named "Spectrum 2".
Manual operation: See "New Channel Setup" on page 83
Selecting the Operating Mode and Application