User Manual

Remote Commands
R&S
®
FPL1000
621User Manual 1178.3370.02 ─ 09
Example:
INST:LIST?
Result for 3 channel setups:
'ADEM','Analog Demod','IQ','IQ
Analyzer','IQ','IQ Analyzer2'
Usage: Query only
Manual operation: See "Changing the Channel Setup Name" on page 64
See "Defining a channel setup" on page 82
See "New Channel Setup" on page 83
Table 11-2: Available channel setup types and default channel setup names
Application <ChannelType> Parameter Default Channel Setup Name*)
Spectrum SANALYZER Spectrum
Analog Modulation Analy-
sis
ADEM Analog Demod
I/Q Analyzer IQ IQ Analyzer
Noise Figure Measure-
ments
NOISE Noise
Vector Signal Analysis
(VSA)
DDEM VSA
Note: the default channel setup name is also listed in the table. If the specified name for a new channel
setup already exists, the default name, extended by a sequential number, is used for the new channel
setup.
INSTrument:REName <ChannelName1>, <ChannelName2>
This command renames a channel setup.
Setting parameters:
<ChannelName1> String containing the name of the channel setup you want to
rename.
<ChannelName2> String containing the new channel setup name.
Note that you cannot assign an existing channel setup name to
a new channel setup; this will cause an error.
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Example:
INST:REN 'IQAnalyzer2','IQAnalyzer3'
Renames the channel setup with the name 'IQAnalyzer2' to
'IQAnalyzer3'.
Usage: Setting only
Manual operation: See "Changing the Channel Setup Name" on page 64