User Manual
Remote Commands
R&S
®
FPL1000
626User Manual 1178.3370.02 ─ 09
//Set marker to maximum in IQ1 and query result
CALCulate:MARKer:MAXimum
CALC:MARK:Y?
//Switch to second IQ channel and retrieve results
INST:SEL 'IQ 2';*WAI
CALCulate:MARKer:MIN
CALC:MARK:Y?
//Switch to first Spectrum channel
INST:SEL 'Spectrum';*WAI
//Query one of the SEM results
CALCulate:MARKer:FUNCtion:POWer:RESult? CPOWer
//Switch to second Spectrum channel
INST:SEL 'Spectrum 2';*WAI
//Query channel power result
CALCulate:MARKer:FUNCtion:POWer:RESult? ACPower
11.6 Configuring and Performing Measurements
The following commands are required to configure measurements in a remote environ-
ment. The tasks for manual operation are described in Chapter 7.2, "Measurements
and Results", on page 96.
● Performing Measurements....................................................................................626
● Configuring Power Measurements........................................................................630
● Measuring Compression Points............................................................................ 634
● Measuring the Channel Power and ACLR............................................................ 638
● Measuring the Carrier-to-Noise Ratio................................................................... 656
● Measuring the Occupied Bandwidth..................................................................... 657
● Measuring the Spectrum Emission Mask..............................................................659
● Measuring Spurious Emissions.............................................................................688
● Analyzing Statistics (APD, CCDF)........................................................................ 701
● Measuring the Time Domain Power...................................................................... 711
● Measuring the Harmonic Distortion.......................................................................720
● Measuring the Third Order Intercept Point............................................................724
● Measuring the AM Modulation Depth....................................................................727
● Remote Commands for EMI Measurements.........................................................729
● List Evaluations.....................................................................................................738
● Measuring the Pulse Power.................................................................................. 742
● Programming Example: Performing a Basic Frequency Sweep........................... 747
11.6.1 Performing Measurements
Useful commands for performing measurements described elsewhere
●
INITiate<n>:ESPectrum on page 661
●
INITiate<n>:SPURious on page 688
Configuring and Performing Measurements