User Manual

Remote Commands
R&S
®
FPL1000
627User Manual 1178.3370.02 ─ 09
Remote commands exclusive for performing measurements:
ABORt.......................................................................................................................... 627
INITiate<n>:CONMeas................................................................................................... 627
INITiate<n>:CONTinuous................................................................................................628
INITiate<n>[:IMMediate]..................................................................................................629
[SENSe:]SWEep:COUNt:CURRent?................................................................................629
ABORt
This command aborts the measurement in the current channel setup and resets the
trigger system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the *OPC? or *WAI command after ABOR and
before the next command.
For details see Chapter 10.1.6.1, "Preventing Overlapping Execution", on page 558.
Note on blocked remote control programs:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S FPL1000 is blocked for further commands. In this case, you must
interrupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S FPL1000 on a parallel channel to clear all currently active remote channels.
Depending on the used interface and protocol, send the following commands:
Visa: viClear()
Now you can send the ABORt command on the remote channel performing the mea-
surement.
Example:
ABOR;:INIT:IMM
Aborts the current measurement and immediately starts a new
one.
Example:
ABOR;*WAI
INIT:IMM
Aborts the current measurement and starts a new one once
abortion has been completed.
Usage: Event
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using ABORt)
or finished in single sweep mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
Suffix:
<n>
.
irrelevant
Configuring and Performing Measurements