User Manual

Remote Commands
R&S
®
FPL1000
637User Manual 1178.3370.02 ─ 09
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example: Enable calculation:
CONF:CME:REF:PNDB ON
Define compression value:
CONF:CME:RES:CVAL 5
Query results:
FETC:CME:PNDB?
Manual operation: See "Compression Points" on page 125
CONFigure:CMEasurement[:STATe] <State>
Enables or disables the compression point measurement.
For details see Chapter 7.2.3, "Power Sweep Measurements", on page 119.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
CONF:CME ON
Manual operation: See "Compression Measurement State" on page 125
FETCh:CMEasurement:P1DB[:RESult]?
FETCh:CMEasurement:P3DB[:RESult]?
FETCh:CMEasurement:PNDB[:RESult]?
Queries the results of the n dB compression point measurement, see CONFigure:
CMEasurement:RESult:PNDB[:STAT] on page 636.
To query the compression value, use CONFigure:CMEasure
ment:RESult:CVALue?.
Example:
FETC:CME:PDNB?
//Result: -3 dBm, 3 dBm
Usage: Query only
Manual operation: See "Compression Points" on page 125
Configuring and Performing Measurements