● Checking Limits.....................................................................................................647
● Performing an ACLR Measurement......................................................................652
● Retrieving and Analyzing Measurement Results.................................................. 653
● Programming Examples for Channel Power Measurements................................ 654
11.6.4.1 Managing Measurement Configurations
The following commands control measurement configurations for ACLR measure-
ments.
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet............................................. 639
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?........................... 639
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:DELete.............................. 640
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE.................................640
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet <Standard>
This command loads a measurement configuration.
The measurement configuration for power measurements consists of weighting filter,
channel bandwidth and spacing, resolution and video bandwidth, detector and sweep
time.
Suffix:
<n>
.
Window
<m> Marker
<sb> irrelevant
Parameters:
<Standard> For more information see Chapter 7.2.4.8, "Reference: Prede-
fined CP/ACLR Standards", on page 148.
If you want to load a customized configuration, the parameter is
a string containing the file name.
Manual operation: See "Predefined Standards" on page 135
See "User Standards" on page 135
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
This command queries all files containing ACLR standards.
Suffix:
<n>
.
Window
<m> Marker
<sb> Multi-SEM: 1 to 3
for all other measurements: irrelevant
Return values:
<Standards> List of standard files.
Usage: Query only
Configuring and Performing Measurements