The following types of information are displayed, if applicable.
Table 5-7: Common settings displayed in the channel bar
SGL The sweep is set to single sweep mode.
Sweep Count The current signal count for measurement tasks that involve a specific number of subse-
quent sweeps
(see "Sweep Count" setting in "Sweep Settings" in the User Manual)
TRG Trigger source
(for details see "Trigger settings" in the User Manual)
●
EXT: External
●
IFP: IF power (+trigger bandwidth)
●
PSE: Power sensor
●
TIM: Time
●
VID: Video
PA The preamplifier is activated.
GAT The frequency sweep is controlled via the "TRIGGER INPUT" connector.
TDF The specified transducer factor is activated.
75 Ω The input impedance of the instrument is set to 75 Ω.
FRQ A frequency offset ≠ 0 Hz is set.
CWSource:
<level>
Internal generator in operation as independant CW source with specified level (requires
option R&S FPL1-B9).
<"NOR" |
"APX">
/Trk.Gen
Internal generator in operation as tracking generator (requires option R&S FPL1-B9).
NOR: the measurements are normalized with the results of the internal generator calibra-
tion
APX (approximation): the measurements are normalized with the results of the internal
generator calibration; however, the measurement settings have been changed since cali-
bration
If neither label is displayed, no calibration has been performed yet or normalization is not
active.
For details, see Chapter 7.3.3.1, "Basics on the Internal Generator", on page 283.
Changing the Channel Setup Name
The channel setups are labeled with their default name. If that name already exists, a
sequential number is added. You can change the name of the channel setup by dou-
ble-tapping the name in the channel bar and entering a new name.
For an overview of default names, see INSTrument:LIST? on page 620.
Remote command:
INSTrument:REName on page 621