User Manual
Remote Commands
R&S
®
FPL1000
652User Manual 1178.3370.02 ─ 09
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage: Query only
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe
This command turns the relative limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
irrelevant
<li> irrelevant
<ch> 1..n
Alternate channel number
Parameters:
<State>
Manual operation: See "Limit Check" on page 140
CALCulate<n>:LIMit<li>:ACPower[:STATe] <State>
This command turns the limit check for ACLR measurements on and off.
Suffix:
<n>
.
irrelevant
<li> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Limit Check" on page 140
11.6.4.6 Performing an ACLR Measurement
The following commands are required to perform an ACLR measurement:
●
CALC:MARK:FUNC:POW:SEL ACP, see CALCulate<n>:MARKer<m>:
FUNCtion:POWer<sb>:SELect on page 632
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe] on page 632
●
INITiate<n>[:IMMediate] on page 629
Configuring and Performing Measurements