User Manual
Remote Commands
R&S
®
FPL1000
659User Manual 1178.3370.02 ─ 09
//Initiate a new measurement and waits until the sweep has finished.
INIT;*WAI
//---------------Retrieving Results-------------
//Return the occupied bandwidth.
CALC:MARK:FUNC:POW:RES? OBW
11.6.7 Measuring the Spectrum Emission Mask
All remote control commands specific to spectrum emission mask measurements are
described here.
See also Chapter 11.6.2, "Configuring Power Measurements", on page 630.
Remote commands exclusive to spectrum emission mask measurements:
● Managing Measurement Configurations............................................................... 659
● Controlling the Measurement................................................................................661
● Configuring a Multi-SEM Measurement................................................................ 661
● Configuring a Sweep List...................................................................................... 662
● Configuring the Reference Range.........................................................................676
● Configuring the Power Classes.............................................................................678
● Configuring the List Evaluation............................................................................. 683
● Performing an SEM Measurement........................................................................685
● Retrieving Results.................................................................................................685
● Example: SEM Measurement............................................................................... 685
11.6.7.1 Managing Measurement Configurations
The following commands control measurement configurations for SEM measurements.
CALCulate<n>:LIMit<li>:ESPectrum<sb>:RESTore............................................................659
[SENSe:]ESPectrum<sb>:PRESet[:STANdard]................................................................. 660
[SENSe:]ESPectrum<sb>:PRESet:RESTore..................................................................... 660
[SENSe:]ESPectrum<sb>:PRESet:STORe....................................................................... 660
CALCulate<n>:LIMit<li>:ESPectrum<sb>:RESTore
This command restores the predefined limit lines for the selected Spectrum Emission
Mask standard.
All modifications made to the predefined limit lines are lost and the factory-set values
are restored.
Suffix:
<n>
.
irrelevant
<li> irrelevant
Configuring and Performing Measurements