Usage: Query only
[SENSe:]ESPectrum<sb>:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
you cannot delete the reference range
a minimum of three ranges is mandatory.
Suffix:
<sb>
.
1 to 3
<ri> Selects the measurement range.
Manual operation: See "Delete Range" on page 179
[SENSe:]ESPectrum<sb>:RANGe<ri>:FILTer:TYPE
<FilterType>
This command selects the filter type for an SEM range.
In case of high speed measurements, the filter has to be identical for all ranges.
The EMI-specific filter types are available if the EMI (R&S FPL1-K54) measurement
option is installed, even if EMI measurement is not active. For details see "Resolution
Bandwidth and Filter Types" on page 253.
Suffix:
<sb>
.
1 to 3
<ri> 1...30
Selects the measurement range.
Parameters:
<FilterType> NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S FPL1-K54) option
Return value for query is always PULS.
MIL
MIL Std (6 dB) - requires EMI (R&S FPL1-K54) option
P5
5 Pole filters
*RST:
NORM
Refer to the datasheet for available filter bandwidths.
Manual operation:
See "Filter Type" on page 175