Power class for which the limit is defined.
Parameters:
<Function> OFF | MAX
Defines the function to be used to determine the relative limit
line start value
MAX
The maximum of the relative and the absolute level is used as
the limit start value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STARt and [SENSe:
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STARt:ABS commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative start value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STARtcommand to
define this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
FUNCtion MAX
For a detailed example see Chapter 11.6.7.10, "Example: SEM
Measurement", on page 685.
Manual operation: See "Rel Limit Start / Stop <n>" on page 177
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP <Level>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit anytime and regardless of the
limit check mode.
Suffix:
<sb>
.
1 to 3
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Relative limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
For a detailed example see Chapter 11.6.7.10, "Example: SEM
Measurement", on page 685.