User Manual
Remote Commands
R&S
®
FPL1000
672User Manual 1178.3370.02 ─ 09
Manual operation: See "Rel Limit Start / Stop <n>" on page 177
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:ABS <Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see "Relative limit line functions" on page 167.
Suffix:
<sb>
.
1 to 3
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STOP:FUNCtion on page 672).
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
For a detailed example see Chapter 11.6.7.10, "Example: SEM
Measurement", on page 685.
Manual operation: See "Rel Limit Start / Stop <n>" on page 177
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:FUNCtion
<Function>
This command enables the use of a function when defining the relative limit for a SEM
range.
Suffix:
<sb>
.
1 to 3
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Function> OFF | MAX
Defines the function to be used to determine the relative limit
line stop value
Configuring and Performing Measurements