User Manual

Remote Commands
R&S
®
FPL1000
673User Manual 1178.3370.02 ─ 09
MAX
The maximum of the relative and the absolute level is used as
the limit stop value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STOP and [SENSe:
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STOP:ABS commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative stop value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STOP command to define
this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
FUNCtion MAX
For a detailed example see Chapter 11.6.7.10, "Example: SEM
Measurement", on page 685.
Manual operation: See "Rel Limit Start / Stop <n>" on page 177
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:STATe <State>
This command selects the limit check mode for all SEM ranges (<range> is irrelevant).
Suffix:
<sb>
.
1 to 3
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<State> ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
AND
Combines the absolute and relative limit. The limit check fails
when both limits are violated.
OR
Combines the absolute and relative limit. The limit check fails
when one of the limits is violated.
*RST:
RELative
Manual operation: See "Limit Check <n>" on page 177
Configuring and Performing Measurements