680User Manual 1178.3370.02 ─ 09
<sb> 1 to 3
Parameters:
<Mode> AUTO
The power class (and thus the limit line) is assigned dynamically
according to the currently measured channel power.
MANUAL
One of the specified power classes is selected manually for the
entire measurement. The selection is made with the
CALCulate<n>:LIMit<li>:ESPectrum<sb>:
PCLass<pc>[:EXCLusive] command.
*RST: AUTO
Example:
CALC:LIM:ESP:MODE AUTO
Activates automatic selection of the limit line.
CALCulate<n>:LIMit<li>:ESPectrum<sb>:VALue <Power>
This command activates the manual limit line selection as and specifies the expected
power as a value. Depending on the entered value, the associated predefined limit
lines is selected.
This command has the same effect as a combination of the CALC:LIM:ESP:MODE
MAN and the CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:
EXCLusive] commands; however, the power class to be used is not defined directly,
but via the expected power. As opposed to CALC:LIM:ESP:MODE AUTO, the power
class is not re-assigned to the input signal power dynamically, but only once when the
command is executed.
Suffix:
<n>
.
irrelevant
<li> irrelevant
<sb> 1 to 3
Parameters:
<Power> integer
Range: -200 to 199
*RST: 0
Example:
CALC:LIM:ESP:VAL 33
Activates manual selection of the limit line and selects the limit
line for P = 33.
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:COUNt <NoPowerClasses>
This command sets the number of power classes to be defined.
This command must be executed before any new power class values can be defined
using CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum and
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum.
Configuring and Performing Measurements