<pc> 1..n
power class
Parameters:
<State> ABSolute | RELative | AND | OR
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
Example:
CALC:LIM:ESP:PCL:LIM ABS
Manual operation: See "Used Power Classes:" on page 182
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum <Level>
This command defines the upper limit of a particular power class.
Note:
●
The last power class always has an upper limit of 200 dBm.
●
The upper limit of a power class must always be the same as the lower limit of the
subsequent power class.
●
The power class must already exist (see CALCulate<n>:LIMit<li>:
ESPectrum<sb>:PCLass<pc>:COUNt on page 680).
Suffix:
<n>
.
irrelevant
<li> irrelevant
<sb> 1 to 3
<pc> 1..n
power class
Parameters:
<Level> Range: -199.9 dBm to 200 dBm
Default unit: dBm
Example:
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40
dBm.
Manual operation: See "PMin/ PMax" on page 183