Example:
LIST:RANG2:INP:GAIN:STAT ON
Manual operation: See "Preamp" on page 206
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue] <Gain>
This command selects the preamplification level for the range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Example:
LIST:RANG2:INP:GAIN 15
[SENSe:]LIST:RANGe<ri>:LIMit:STARt <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level> Absolute limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: 13
Default unit: dBm
Example:
LIST:RANG2:LIM:STAR 200
Manual operation: See "Abs Limit Start/ Abs Limit Stop" on page 207
[SENSe:]LIST:RANGe<ri>:LIMit:STATe <State>
This command turns the limit check for all spurious emission measurement ranges on
and off.
Suffix:
<ri>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
LIST:RANG2:LIM:STAT ON
Manual operation: See "Limit Check" on page 207
[SENSe:]LIST:RANGe<ri>:LIMit:STOP <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Configuring and Performing Measurements